Introduced more than 30 years ago, the COBE 2991 cell processor has set the industry standard as a reliable laboratory tool for a variety of cell processing applications. It earned its enviable reputation for reliability, ease of use, instrument consistency, disposable set performance and multiapplication flexibility thanks to its many devoted users.
The COBE 2991 cell processor has been used for these applications:
Watch a cell processing procedure from start to finish
Learn more nowThe key features of the COBE 2991 cell processor are its centrifuge and its flexible diaphragm. When the centrifuge is spinning, the flexible diaphragm located inside the centrifuge bowl can be inflated with hydraulic fluid. As the flexible diaphragm inflates, it presses against the bottom of the cell processing bag. This gives the COBE 2991 cell processor the ability to express fluids and/or cells for removal or collection during centrifugation. The clear centrifuge cover allows you to visually observe cell separation and fluid expression.
Width | 46 cm/18 in |
---|---|
Depth | 77 cm/30 in |
Counter height | 96 cm/37.5 in |
Top of control box | 155 cm/61 in |
Weight | 195 kg/430 lb |
Heat output | 479 kcal/hr 1,900 BTU/hr |
Air flow | 70 CFM (2 m3) See operating clearance. |
Operating clearance | A minimum of 38 cm (15 in) is required on either the right or the left side. A 5.1 cm (2 in) clearance must be maintained at the rear for air circulation.
Clearance at the front is 76 cm (30 in). |
Service clearance | A minimum of 76 cm (30 in) is required on all sides for service. Device is equipped with fixed casters in the rear and swivel casters in the front to permit moving the machine for servicing. Casters have a locking device to prevent movement during operation. |